Removing the monitor board and spo2 modules, Removing the monitor board and spo, Modules – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 86

Advertising
background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

80

9650-001356-01 Rev. D

Removing the Monitor Board and SpO

2

Modules

Tool Required

# 1 Phillips screwdriver with torque capability

Wooden stick

Setup

This procedure should be conducted on a static mat.

Wear a wrist strap to ground during the procedure.

Complete these procedures in the following order before performing this procedure:

1.

“Removing the Handle” on page 63

2.

“Removing the Printer Module” on page 64

3.

“Disassembling the Main Unit” on page 66

4.

“Removing the Display Module” on page 72

5.

“Removing the EtCO2 and NIBP Modules” on page 76

To reinstall the monitor board and/or SpO

2

modules, reverse the steps.

Monitor Board and/or SpO

2

Disassembly

1

Remove 4 screws on the right
side of the unit in the
illustrated areas.

Note: During reassembly,
torque screws to 6 in-lbs.

Advertising