0 treatment report test, Treatment report test – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 35

Advertising
background image

Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

29

13.0 Treatment Report Test

Equipment

Fluke Impulse 7000 DP Analyzer (or equivalent)

Do this...

Observe this...

Pass/Fail

13.1 Connect the multifunction cable (MFC) to the

Analyzer, and turn on the Analyzer.

13.2 Turn on the X Series unit, and press the

following quick access keys:

a)

b)

c)

The Clear Log window
displays.

13.3 Select Yes.

The unit displays a clearing
the log message.

o

o

13.4 Press the CHARGE button, select 200 joules,

then press the CHARGE button again.

The unit starts to charge, a
charging tone sounds, and the
Shock button illuminates.

o

o

13.5 Press the Shock button.

The unit discharges the shock.

o

o

13.6 Wait 20 seconds, press the Treatment quick

access key(

), and select the first code

marker.

13.7 Wait 15 seconds and turn off the unit, then wait

2 minutes and turn on the unit.

13.8 On the X Series unit, press the following quick

access keys:

a)

b)

c)

The Treatment Summary
window displays.

o

o

13.9 Highlight the second event and press Select.

Verify that a green check mark
displays next to the file.

o

o

13.10 Select Print Treatment Summary.

Verify that the recorder prints
out the defib event and the
code marker snapshot.

o

o

13.11 Turn off the X Series unit.

LOG

R

LOG

Advertising