Removing the EtCO2 and NIBP Modules
9650-001356-01 Rev. D
X Series Service Manual
77
2
Remove four (4) screws
around the EtCO
2
unit.
Note: During reassembly,
torque screws to 6 in-lbs.
3
Disconnect the EtCO
2
sensor
from side panel by using a
metal pick and releasing the
clips on both sides.
4
Disconnect the exhaust tube
connector from the air vent at
area shown.