ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 83

Advertising
background image

Removing the EtCO2 and NIBP Modules

9650-001356-01 Rev. D

X Series Service Manual

77

2

Remove four (4) screws
around the EtCO

2

unit.

Note: During reassembly,
torque screws to 6 in-lbs.

3

Disconnect the EtCO

2

sensor

from side panel by using a
metal pick and releasing the
clips on both sides.

4

Disconnect the exhaust tube
connector from the air vent at
area shown.