Spo2 test (for spo, 0 spo2 test (for spo, Feature without spco and spmet) – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 36: 0 spo, Test (for spo

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

30

9650-001356-01 Rev. D

14.0 SpO

2

Test (for SpO

2

feature without SpCO and SpMet)

Note:

Perform this test if your unit has SpO

2

only. See the following procedure to test units

with SpO

2

, SpCO, and SpMet.

Equipment

Fluke Index 2 SpO

2

Simulator (or equivalent), Masimo

®

Rainbow

®

SpO

2

finger probe

Test Setup

Disconnect any ECG cable or IBP probes from the X Series unit.

Do This...

Observe This...

Pass/Fail

14.1 Power on the X Series unit without the

SpO

2

sensor attached.

Verify that the

SpO

2

numeric

displays INITIALIZING, then
goes blank.

14.2 Connect the

SpO

2

sensor to the X Series unit. Verify that

SpO

2

numeric

displays SEARCHING.

14.3 Set the simulator to 70%

SpO

2

and 60 PPM.

Attach the

SpO

2

sensor to the simulator.

Verify that after a brief period of
initialization, the

SpO

2

numeric

displays an

SpO

2

value between

67% and 73%.

Verify that the displayed pulse
rate reads 60 +/- 3 bpm.

o

o

o

o

14.4 Set the simulator to 80%

SpO

2

and 60 PPM.

Verify that after a brief period of
initialization, the

SpO

2

numeric

displays an

SpO

2

value between

77% and 83%.

Verify that the displayed pulse
rate reads 60 +/- 3 bpm.

o

o

o

o

14.5 Set the simulator to 90%

SpO

2

and 60 PPM.

Verify that after a brief period of
initialization, the

SpO

2

numeric

displays an

SpO

2

value between

87% and 93%.

Verify that the displayed pulse
rate reads 60 +/- 3 bpm.

o

o

o

o

14.6 Disconnect the

SpO

2

simulator.

Advertising