C
HAPTER
3
D
ISASSEMBLY
P
ROCEDURES
66
9650-001356-01 Rev. D
Disassembling the Main Unit
Tool Required
•
# 1 Phillips screwdriver with torque capability
•
Wooden stick
•
Kapton tape
Setup
•
This procedure should be conducted on a static mat.
•
Wear a wrist strap to ground during the procedure.
•
Complete these procedures in the following order before performing this procedure:
1.
“Removing the Handle” on page 63
2.
“Removing the Printer Module” on page 64
To reassemble the main unit, reverse the steps.
Main Unit Disassembly
1
Place tape on both sides of
the unit. (This will temporary
hold the sections of the unit
together).