Disassembling the main unit – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 72

Advertising
background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

66

9650-001356-01 Rev. D

Disassembling the Main Unit

Tool Required

# 1 Phillips screwdriver with torque capability

Wooden stick

Kapton tape

Setup

This procedure should be conducted on a static mat.

Wear a wrist strap to ground during the procedure.

Complete these procedures in the following order before performing this procedure:

1.

“Removing the Handle” on page 63

2.

“Removing the Printer Module” on page 64

To reassemble the main unit, reverse the steps.

Main Unit Disassembly

1

Place tape on both sides of
the unit. (This will temporary
hold the sections of the unit
together).

2

Gently place unit screen so
the display is facing down.

Advertising