0 etco2 calibration check, Etco2 calibration check, 0 etco – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 40: Calibration check

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

34

9650-001356-01 Rev. D

17.0 EtCO

2

Calibration Check

Equipment

5% CO

2

calibration gas cylinder with CO

2

sample line (kit available from Scott Medical

Products, PN 4653ORF-BD)

Note:

If you have previously run the CO2 Calibration procedure, start at step 3.

Do This...

Observe This...

Pass/Fail

17.1

On the X Series unit, select
Supervisor>Service>Diagnostic>CO

2

>

Calibration to enter the CO

2

Calibration

control panel.

17.2

Connect the X Series unit to the 5% CO

2

calibration gas cylinder using the sample
line included with the calibration kit.

17.3

From the CO

2

Calibration control panel,

select Verify Calibration.

17.4

Activate the calibration gas flow by
pressing the top valve on the calibration gas
canister.

17.5

Continue to apply the calibration gas until
the status in the CO

2

Calibration control

panel displays: CALIBRATION GAS CAN
BE REMOVED
.

o

o

17.6

Verify that the unit displays:
VERIFICATION DONE - OK.

o

o

17.7

Verify that the measured gas reading is
between 4.8 and 5.2 (Vol%).

o

o

17.8

Exit out of calibration by pressing the Back

arrow key (

) three times or pressing the

Display/Home button (

).

Advertising