Maintenance tests, 0 physical inspection of the unit, Physical inspection of the unit – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 14

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

8

9650-001356-01 Rev. D

Maintenance Tests

1.0 Physical Inspection of the Unit

Observe this...

Pass/Fail

1.1

Housing

Is the unit clean and undamaged?

o

o

1.2 Does the unit show signs of excessive wear?

o

o

1.3 Does the handle work properly?

o

o

1.4 Does the recorder drawer open and close properly?

o

o

1.5 Are input connectors clean and undamaged?

o

o

1.6 Are there any cracks in the housing?

o

o

1.7 Does the front panel have any damage or cracks?

o

o

1.8 Are there any loose housing parts?

o

o

1.9 Paddles (if applicable)

Do the adult and pedi plates have major scratches or show signs of damage?

o

o

1.10 Do the adult shoes slide on and off easily to expose the covered pedi plates?

o

o

1.11 Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)

o

o

1.12 Cables

Are all cables free of cracks, cuts, exposed or broken wires?

o

o

1.13 Are all bend/strain reliefs undamaged and free of excessive cable wear?

o

o

1.14 Battery

Is the X Series battery fully charged?

o

o

1.15 Is the battery seated in the battery well correctly?

o

o

1.16 Is battery housing free of cracks and damage?

o

o

1.17 Are connectors free of damage and excessive wear?

o

o