Maintenance Tests
9650-001356-01 Rev. D
ZOLL X Series Service Manual
39
20.25 Once complete, use the navigation keys to
select Update NIBP Cal Check Date to
save the date of the most recent
calibration check.
Note: This will reset the date. Nothing
displays on the screen.
20.26 Press
twice and select
Diagnostic>NIBP>Calibration Info.
Verify that the last NIBP Cal date is
today’s date.
20.27 Exit out of the Supervisor menu.
Do This...
Observe This...
Pass/Fail