ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 45

Advertising
background image

Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

39

20.25 Once complete, use the navigation keys to

select Update NIBP Cal Check Date to
save the date of the most recent
calibration check.
Note: This will reset the date. Nothing
displays on the screen.

20.26 Press

twice and select

Diagnostic>NIBP>Calibration Info.
Verify that the last NIBP Cal date is
today’s date.

20.27 Exit out of the Supervisor menu.

Do This...

Observe This...

Pass/Fail

Advertising