ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 24

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

18

9650-001356-01 Rev. D

5.12 Set simulator to 160 BPM or higher.

Heart Rate Value is highlighted and
the alarm tone sounds. The HR/PR
HIGH ALARM
message displays.

o

o

5.13 On the X Series unit, press

the

Alarm Silence/Reset button

(

)

.

Alarm is suspended for 90 seconds.
The alarm suspension timer appears
in black.

o

o

Do this...

Observe this...

Pass/Fail