Removing the dock connector – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 102

Advertising
background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

96

9650-001356-01 Rev. D

Removing the Dock Connector

Tools Required

#1 Phillips screwdriver with torque capability

Setup

This procedure should be conducted on a static mat.

Wear a wrist strap to ground during the procedure.

Complete these procedures in the following order before performing this procedure:

1.

“Removing the Handle” on page 63

2.

“Removing the Printer Module” on page 64

3.

“Disassembling the Main Unit” on page 66

4.

“Removing the Defibrillator Boards and Capacitor Bank Modules” on page 88

Dock Connector

1

Remove 4 screws and
washers from the back panel
that are securing the dock
connector to the housing.

Note: During reassembly,
torque screws to 6 in-lbs.

2

Disconnect black connector
from the dock connector
board.

Advertising