Removing the display module – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 78

C
HAPTER
3
D
ISASSEMBLY
P
ROCEDURES
72
9650-001356-01 Rev. D
Removing the Display Module
Tools Required
•
#1 Phillips screwdriver with torque capability
•
3M copper adhesive tape
•
6mm socket wrench
•
Mid-Frame LCD Cable Routing Assembly Fixture, 9100-3028-AF (Optional)
Setup
•
This procedure should be conducted on a static mat.
•
Wear a wrist strap to ground during the procedure.
•
Complete these procedures in the following order before performing this procedure:
1.
“Removing the Handle” on page 63
2.
“Removing the Printer Module” on page 64
3.
“Disassembling the Main Unit” on page 66
4.
“Removing the Front Enclosure Module” on page 70
To reinstall the display module, reverse the steps.
Display Disassembly
1
Remove the grounding screw
and washer for EtCO
2
in the
area illustrated.