C
HAPTER
3
D
ISASSEMBLY
P
ROCEDURES
86
9650-001356-01 Rev. D
2
Remove the 9-pin GPS and 2
antenna cables (gray cable
J7 and black cable J6) from
the CP carrier board.
3
Turn board over, and remove
the 4 screws securing the CP
board to the monitor board.
Note: During reassembly,
torque screws to 6 in-lbs.
4
Only lift the CP carrier board
(not the metal heat spreader)
upwards at location J60 on
the monitor board.
Note: During reassembly,
align connectors on board.