0 leads test, Leads test – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 21

Maintenance Tests
9650-001356-01 Rev. D
ZOLL X Series Service Manual
15
3.0 Leads Test
Equipment
3-lead, 5-lead, and 12-lead cables, Fluke Impulse 7000DP Analyzer
Note:
Test each cable separately.
Test Setup
Connect the lead wires appropriate for each test to the
Fluke Impulse 7000DP Analyzer
or
equivalent, then connect the ECG cable to the X Series unit.
Do this...
Observe this...
Pass/Fail
3.1 Turn on the X Series unit.
Note: Make sure the lead is not
displayed.
No ECG LEAD FAULT message
displayed.
o
o
3.2 Disconnect one lead from the
simulator.
The ECG LEAD FAULT message displays
within 3 seconds (if configured).
o
o
3.3 Reconnect the lead. Repeat step 2
with the remaining leads.
Wait for ECG LEAD FAULT message to
clear from the display (if configured).
o
o
3.4 Repeat steps 2 and 3 for 5 lead and 12
lead cables.
Note: If heart rate alarm sounds, on the
X Series unit:
Press the More quick access key (
) key
to access the second set of quick access
keys.
Press the Alarm quick access key (
).
Press the Alarm Cancel quick access key
(
).
o
o