0 nibp leak test, Nibp leak test – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 46

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

40

9650-001356-01 Rev. D

21.0 NIBP Leak Test

Equipment

Dual lumen NIBP hose, stop watch, NIBP simulator

Do This...

Observe This...

Pass/Fail

21.1 Turn on the X Series unit.

Note: If the unit is in AED mode, press
the Manual quick access key to go into
manual mode.

21.2 Connect a dual lumen hose from the NIBP

simulator to the NIBP port of the X
Series.

21.3 On the X Series unit, do the following to

enter the NIBP Test control panel:

Press the More (

) then the Setup

(

) quick access keys.

Use the navigation keys to select
Supervisor>Service>Device
Tests
>NIBP Test.

Note: You need a supervisor passcode to
enter the Supervisor menu.

21.4 From the X Series NIBP Test control

panel, use the navigation keys to select:

Close Valve

Turn Pump On

21.5 Monitor the pressure reported by the

external pressure standard.

21.6 Once the pressure reaches the target

pressure of 250 +/- 5 mmHg, use the
navigation keys to select Turn Pump Off.

21.7 Allow the system to stabilize for at least

10 seconds.

21.8 Select Mark P1 to mark the starting

pressure. Wait 15 seconds, then select
Mark P2.

21.9 From the NIBP Test control panel, select

Open Valve to release the pressure.

21.10 Calculate the leak rate: L = P1 - P2.

Reading should be=<5 mmHg

o

o

21.11 From the NIBP Test control panel, select

Clear.

Advertising