15.5 Set the simulator to 95%
SpO
2
and 230
BPM.
Verify that after a brief period of
initialization,
The
SpO
2
numeric displays an
SpO
2
value between 91.3% and
98.8%.
The
SpCO
numeric displays an
SpCO
value between -.5 and 7.5.
The
SpMet
numeric displays an
SpMet
value between -.5 and 3.5.
Verify that the displayed pulse
rate reads between 222.7 and 236
BPM.
o
o
o
o
o
o
o
o
o
15.6 Set the simulator to 95%
SpO
2
and 50
BPM.
Verify that after a brief period of
initialization,
The
SpO
2
numeric displays an
SpO
2
value between 91.3% and
98.8%.
The
SpCO
numeric displays an
SpCO
value between -.5 and 7.5.
The
SpMet
numeric displays an
SpMet
value between -.5 and 3.5.
Verify that the displayed pulse
rate reads between 44.5 and 55.5
BPM.
o
o
o
o
o
o
o
o
o
15.7 Disconnect the
SpO
2
simulator.
Do This...
Observe This...
Pass/Fail/N/A