Removing the front enclosure module – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 76

Advertising
background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

70

9650-001356-01 Rev. D

Removing the Front Enclosure Module

Tools Required

#2 Phillips screwdriver with torque capability

Wooden stick

Setup

This procedure should be conducted on a static mat.

Wear a wrist strap to ground during the procedure.

Complete these procedures in the following order before performing this procedure:

1.

“Removing the Handle” on page 63

2.

“Removing the Printer Module” on page 64

3.

“Disassembling the Main Unit” on page 66

To reinstall the front enclosure module, reverse the steps.

Front Enclosure Disassembly

1

Remove tape from both sides
of the unit.

2

Separate front enclosure
assembly.

3

Pivot top up with cables still
connected.

Advertising