ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 26

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

20

9650-001356-01 Rev. D

6.5

On the X Series unit, do the
following:

Press the More (

) then the

Setup (

) quick access keys.

Use the navigation keys to
select Supervisor>Display/
Configuration>Device Info
and select Restore All Factory
Defaults
.

Note: You need a supervisor
passcode to enter the Supervisor
menu. Also the unit will reboot
during the of restoring factory
defaults.

6.6 Press

the

I Measure switch on the

Battery Current Test Fixture.

Observe the current measurement on
the DMM, and verify that the reading
is between 0.47A-1.25A.

o

o

Do this...

Observe this...

Pass/Fail