Safety precautions, Modules, Safety precautions modules – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 68

Advertising
background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

62

9650-001356-01 Rev. D

Kapton tape

3M copper adhesive tape, or equivalent

Large diagonal cutters

Safety Precautions

WARNING!

SHOCK HAZARD!

Caution

TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY

BEFORE YOU CONDUCT DEFIBRILLATOR TESTS OR REPAIRS.

Only properly trained technicians should service the unit.

The unit can contain deadly voltages even if the unit is turned off.

Make sure to discharge the unit before working with it.

Make sure you take the necessary precautions when working with static sensitive units. For
example, you must wear a conductive wrist strap (which touches your skin) connected to a
grounding mat and to the earth ground. You must remove the wrist strap when you discharge
high voltage or when you are working on energized equipment.

Modules

The X Series unit contains the following ten (10) modules or assemblies:

Printer

Front Enclosure

Display

NIBP

EtCO

2

(Oridion

®

)

Monitor Board

Capacitor Bank

Defibrillator Board

SpO

2

CP Carrier Module

These modules can be purchased as replacement parts from the ZOLL Technical Support
Department. See Chapter 4, "Replacement Parts" for information on ordering replacement
parts.

Advertising