ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 44

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

38

9650-001356-01 Rev. D

20.10 From the NIBP Test control panel, select

Open Valve to release the pressure.

20.11 From the X Series NIBP Test control

panel, use the navigation keys to select:

Close Valve

Turn Pump On

20.12 Monitor the pressure reported by the

external pressure standard.

20.13 Once the pressure reaches the target

pressure of 150 +/- 5 mmHg, use the
navigation keys to select Turn Pump Off.

20.14 Allow the system to stabilize for at least

10 seconds.

20.15 Verify that the primary transducer agrees

with the reading on the NIBP simulator.

Reading should be within
3mmHg.

o

o

20.16 Verify that the safety transducer agrees

with the reading on the NIBP simulator.

Reading should be within
4mmHg.

o

o

20.17 From the NIBP Test control panel, select

Open Valve to release the pressure.

20.18 From the X Series NIBP Test control

panel, use the navigation keys to select:

Close Valve

Turn Pump On

20.19 Monitor the pressure reported by the

external pressure standard.

20.20 Once the pressure reaches the target

pressure of 250 +/- 5 mmHg, use the
navigation keys to select Turn Pump Off.

20.21 Allow the system to stabilize for at least

10 seconds.

20.22 Verify that the primary transducer agrees

with the reading on the NIBP simulator.

Reading should be within
3mmHg.

o

o

20.23 Verify that the safety transducer agrees

with the reading on the NIBP simulator.

Reading should be within
4mmHg.

o

o

20.24 From the NIBP Test control panel, select

Open Valve to release the pressure.

Do This...

Observe This...

Pass/Fail