0 power supply test, Power supply test – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 25

Advertising
background image

Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

19

6.0 Power Supply Test

Equipment

Battery Current Test Fixture (P/N 9100-3055-TF), Power Supply, Digital Multi-meter.

Test Setup

1. Set the Power Supply to 12V VDC 10 A.

2.

Turn off the Power Supply output.

3.

Connect the positive (PS+) post of the Power Supply to the positive (PS+) post (PS Red) of
the Battery Current Test Fixture.

4.

Connect the negative (PS-) post of the Power Supply to the negative (PS-) post (PS Black) of
the Battery Current Test Fixture.

5.

With the Battery Current Test Fixture out of the battery well, turn on the Power Supply
output.

6.

Set the Digital Multi-meter to measure DC I DC Amps.

7.

Set the rear/front switch on the Digital Multi-meter to rear.

8.

Connect the Digital Multi-meter (I+) terminal of the Battery Current Test Fixture to the red
terminal (I) on the back of the Digital Multi-meter.

9.

Connect Digital Multi-meter (I-) terminal of the Battery Current Test Fixture to the black
terminal (LO) on the back of the Digital Multi-meter.

Do this...

Observe this...

Pass/Fail

6.1

Install the Battery Current Test
Fixture into battery well and wait
about five seconds.

6.2

Make sure the X Series unit is
turned off.

6.3 Press

the

I Measure switch on the

Battery Current Test Fixture.

Verify that the reading on the digital
multi-meter is between
0.583mA-0.770mA.

o

o

6.4

Turn on the X Series unit, and wait
about ten seconds until the power
has stabilized.