0 advisory test, Advisory test – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 31

Maintenance Tests
9650-001356-01 Rev. D
ZOLL X Series Service Manual
25
9.0 Advisory Test
Equipment
Fluke Impulse 7000 DP Analyzer (or equivalent), ECG simulator
Do This...
Observe This...
Pass/Fail
9.1
Connect the multifunction cable (MFC) to
the Analyzer.
9.2
Turn on the X Series unit and connect the
pads.
The unit displays Pads on the top
left of the screen.
o
o
9.3
On the Analyzer, select VF (ventricular
fibrillation), then press the Analyze button.
The X Series unit displays:
ANALYZING
SHOCK ADVISED
o
o
9.4
On the X Series unit, press the CHARGE
button to activate, then press the CHARGE
button again.
The unit charges to the selected
energy level.
o
o
9.5
On the X Series unit, press the Shock button. Verify that the shock count is
displayed at the top of the screen.
o
o
9.6
On the Analyzer, select NSR (normal sinus
rhythm), then press the Analyze button.
The X Series unit displays:
ANALYZING
NO SHOCK ADVISED
o
o