0 etco, Calibration (required for annual calibration) – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 39

Maintenance Tests
9650-001356-01 Rev. D
ZOLL X Series Service Manual
33
16.0 EtCO
2
Calibration (required for annual calibration)
Equipment
5% CO
2
calibration gas cylinder with CO
2
sample line (kit available from Scott Medical
Products, PN 4653ORF-BD)
Do This...
Observe This...
Pass/Fail
16.1
Turn on the X Series unit.
16.2
On the X Series unit, select
Supervisor>Service>Diagnostic>CO
2
>
Calibration to enter the CO
2
Calibration
control panel.
16.3
Connect the X Series unit to the 5% CO
2
calibration gas cylinder using the sample
line included with the calibration kit.
16.4
From the CO
2
Calibration Menu, select
Calibration.
16.5
Activate the calibration gas flow by
pressing the top valve on the calibration gas
canister.
16.6 Continue to apply the calibration gas until the
status in the CO
2
Calibration control panel
displays: CALIBRATION GAS CAN BE
REMOVED.
•
When the calibration is complete, the
status in the CO
2
Calibration control panel
displays: CALIBRATION DONE-OK
The calibration date will be updated upon
exiting the CO
2
Test control panel.
o
o
o
o