X Series Error Messages

9650-001356-01 Rev. D

X Series Service Manual

59

TEMP DEVICE FAILURE

Temp software detected a
problem with the
temperature data that is
most likely a hardware issue.

Replace the main board.

TEMP PROBE1 (or
PROBE2) SHORTED

An A/D value was detected
that indicated the probe was
shorted.

Replace the temperature
probe or adapter cable. If
problem persists, replace
the right side panel.

TRANSMISSION FAILED

The communication
operation failed. This
message displays until a
new operation is started.

To correct the problem,
verify the following:

Wireless communication
is enabled on the X
Series unit.

The Wi-Fi settings are
correct in the
Communications setup
menu.

The ZOLL server is
configured correctly.

Your cellular device is
configured correctly.

The X Series unit is
within range of the
wireless server.

WAN CONNECTION
UNAVAILABLE

The X Series unit was
unable to establish a WAN
connection.

To correct the problem,
verify the following:

Wireless communication
is enabled on the X
Series unit.

The expected means of
coummunication is
enabled (for instance, if
using a cell modem,
make sure the USB cell
modem is set to
Enabled).

(Bluetooth) It is paired
with the X Series unit.

(Wi-Fi) The correct Wi-Fi
access point profile is
selected.

(Cellular modem) It is
properly connected to
the X Series unit and
that the USB icon on the
display is illuminated.

Error Message

Explanation

Technical Action