Removing the printer module – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 70

Advertising
background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

64

9650-001356-01 Rev. D

Removing the Printer Module

Tools Required

#1 Phillips screwdriver with torque capability

Setup

This procedure should be conducted on a static mat.

Wear a wrist strap to ground during the procedure.

To reinstall the printer module, reverse the steps.

Printer Disassembly

1

Place the unit on the mat in a
horizontal position.

2

Open the printer housing and
remove the printer paper.

3

Place the unit screen down
and remove the three (3)
screws outside the printer
module in the illustrated
areas.

Note: During reassembly,
torque screws to 6 in-lbs
.

Advertising