Setup items – Yokogawa Value Series FX1000 User Manual

Page 103

Advertising
background image

3-30

IM 04L21B01-01EN

Setup Items

• Input Range

First-CH, Last-CH

Select the target channels.

Range > Mode

Setting

Name

LogType1 Log input (Nonlinear log input is possible through the use of calibration correction.)
LogType2 Log linear input or pseudo log input.

Range > Range
Set the range to

20mV, 60mV, 200mV, 2V, 6V, 20V, 50V, or 1V.

Range > Span_L and Span_U

Set the settable range for the Range setting. You have to set these settings so that

Span_L is less than Span_U.

Range > Scale_L, Scale_U

Set the values by specifying the mantissas and exponents. Set the mantissas with

two decimal places.

When Mode is LogType1

Set the settings so that

Scale_L is less than Scale_U. Set the mantissas to values

in the range of 1.00 to 9.99. Set the mantissas using three digits.

Mantissa of Scale_L Setting Range

1.00

1.00E–15 to 1.00E+15. However, the difference between the

exponents must be at least 1.
The scale width must be 15 decades or less.

Scale_L

Scale_U

Notes

Example

1.00E+01

1.00E+02

Examples of

invalid settings

1.00E+01

2.00E+01

The difference

between the

exponents is less

than 1.

1.00E–01

1.00E+15

The scale width

is greater than 15

decades.

A value other than 1.00 1.01E–15 to 1.00E+15. However, the difference between the

exponents must be at least 2.
The scale width must be 15 decades or less.

Scale_L

Scale_U

Notes

Examples

2.00E+01

1.00E+03

2.00E+00

5.00E+14

Example of invalid

settings

2.00E+01

7.00E+02

The difference

between the

exponents is less

than 2.

2.00E–01

1.00E+15

The scale width

is greater than 15

decades.

2.00E+03

2.00E+15

The upper limit

is greater than

1.00E+15.

3.13 Using the Log Scale to Perform Measurements (/LG1 Option)

Advertising