App index standard performance – Yokogawa Value Series FX1000 User Manual

Page 284

Advertising
background image

13-19

IM 04L21B01-01EN

Specifications

3

2

1

4

5

6

7

8

9

10

11

12

13

14

App

Index

Standard Performance

Item

Specifications

Measurement/display accuracy

Standard operating conditions:

Temperature:

23 ± 2°C

Humidity:

55 ± 10%RH

Power supply voltage:

90 to 132 or 180 to 250 VAC

Power supply frequency: 50/60 Hz ± 1%

Warm-up time:

At least 30 minutes.

Other ambient conditions such as vibration should not adversely affect the operation.

Input Type

Range

Measurement Accuracy (Digital display)

Digital Display

Max. Resolution

DC voltage

20 mV

±(0.05% of rdg + 12 digits)

1 μV

60 mV

±(0.05% of rdg + 3 digits)

10 μV

200 mV

10 μV

1 V

100 μV

2 V

±(0.05% of rdg + 12 digits)

100 μV

1 to 5 V

±(0.05% of rdg + 3 digits)

1 mV

6 V

1 mV

20 V

1 mV

50 V

10 mV

Thermocouple
(Not including

the accuracy of

reference junction

compensation; when

the burnout detection

function is off.)

R

±(0.15% of rdg + 1°C)
R, S

0 to 100°C: ±3.7°C,

100 to 300°C: ±1.5°C

B

400 to 600°C: ±2°C; accuracy not guaranteed for

temperatures less than 400°C

0.1°C

S

B

K

±(0.15% of rdg + 0.7°C)
–200 to –100°C: ±(0.15% of rdg + 1°C)

E

±(0.15% of rdg + 0.5°C)
–200 to –100°C: ±(0.15% of rdg + 0.7°C)

J

T

N

±(0.15% of rdg + 0.7°C)
–200 to 0°C: ±(0.35% of rdg + 0.7°C)
Accuracy not guaranteed for temperatures less than –200°C.

W

±(0.15% of rdg + 1°C)

L

±(0.15% of rdg + 0.5°C)
–200 to –100°C: ±(0.15% of rdg + 0.7°C)

U

WRe

±(0.2% of rdg + 2.5°C)
0 to 200°C: ±4.0°C

RTD

Pt100

±(0.15% of rdg + 0.3°C)

JPt100

DI

Voltage

Threshold level (Vth=2.4 V) accuracy ± 0.1 V

Contact

With parallel capacitance of 0.01 μF or less, 1 kΩ or less: 1 (ON). 100 kΩ or more: 0 (OFF)

Measuring accuracy in case of scaling

Accuracy during scaling (digits) = measurement accuracy (digits) × multiplier + 2 digits (rounded up)

* Fractions rounded up

where the multiplier = scaling span (digits)/measuring span (digits).

Example For 1-5 V range (A/D integration time is 16.7 ms or more), measurement span of 1.000 to

5.000 V, and scaling span of 0.000 to 2.000

The measuring accuracy for 5 V input is as follows.

Measuring accuracy (1-5 Vrange) = ±(0.05% × 5 V + 3 digits) = ±(0.0025 V [3 digits] + 3

digits) = ±6 digits

Multiplier = {2000 digits (0.000 to 2.000)}/4000 digits (1.000 to 5.000) = 0.5

Thus, accuracy during scaling = ±(6 × 0.5 + 2) digits = 5 digits (rounded up)

Reference junction compensation accuracy

When measuring temperature greater than or equal to 0 °C and when input terminal temperature is

balanced

Type R, S, W, WRe: ±1.0°C

Type K, J, E, T, N, L, and U: ±0.5°C

Type B: Internal reference compensation is fixed to 0°C

13.6 General Specifications

Advertising