Bar graph – Yokogawa Value Series FX1000 User Manual

Page 19

Advertising
background image

1-10

IM 04L21B01-01EN

Status Display Section

The following information is displayed in the status display section during operation mode
or setting mode.

Memory sampling status

CF card icon (on FXs that have a CF card slot)

Waiting.

Light blue icon: CF card in the slot is not

recognized. Remove and reset it.
CF card error.

Carry out the procedure below to reset the CF

card icon to normal.

• Remove the CF card, and then reinsert it.

• Replace the CF card with a normal one.

• Use the FX to format the CF card (the CF

card data will be deleted).

The green level display indicates the amount of

CF card used. If Media FIFO* is not enabled and

the free space on the CF card falls below 10%,

the level indicator changes to red.

* See section 1.4 in the FX1000 User’s Manual.

Memory sampling icon

Memory sampling

stopped

Memory sampling

in progress

Computation icon (/M1, /PM1, and /PWR1 options)

(Red)

(Green)

CF card is being accessed.

The status assigned to the status output

(/F1 option) is occurring.

Keys are locked.

E-mail transmission (/C7 option) is enabled.

Status icon

Data type

DISP: Display data

EVENT: Event data

Memory sampling progress

Displays the progress using a green bar graph. The frame indicates the file

save interval (display data) or the data length (event data).

Error in internal memory.

Contact your nearest YOKOGAWA dealer.

Displays the remaining memory sampling time for the left bar graph.

Date and time

Displayed in yellow while the time is being corrected.

Display name or group name

For all channel display on the trend display, “All” is

displayed.

When using the batch function

If the “batch number-lot number” exceeds 20 characters,

the “date and time” position is used to display the “batch

number-lot number.”

Batch name and the display

name are shown alternately.
Date and time

Name of the user logged in

Batch name, the display

name, and date and time are

shown alternately.

When using the login function

Name of the user logged in

When using the login and batch functions

Date and time and the

display name are shown

alternately.

Displayed when any alarm is activated.
Blinks when there are alarms that are
occurring but have not been acknowledged.

All alarms have been released after they
have occurred, but there are alarms that
have not been acknowledged.

Alarm icon

White icon: Computation started

Yellow icon: Computation data dropout occurred

Red icon: Error in the power measurement

section

Bar Graph

When event data recording is set to pretrigger, the FX will start recording pretrigger
data after you press the START key. “Waiting” appears in the bar graph. At this time,
the progress bar will turn orange. After the pretrigger time elapses, the length of the bar
fixed at that point. However, the relevant data is updated until the trigger condition is met.
When the trigger condition is met, the bar turns green, and data is recorded after the data
in the pretrigger section.

1.3 Display

Advertising