5 options, 5 options -10, Alarm output relay (/a1, /a2, /a3, and /a4a) – Yokogawa Value Series FX1000 User Manual

Page 275: Ethernet communication interface (/c7)

Advertising
background image

13-10

IM 04L21B01-01EN

13.5 Options

Alarm Output Relay (/A1, /A2, /A3, and /A4A)

Item

Specifications

Action

Outputs relay contact signals from the terminals on the rear panel when alarms occur.

Number of outputs

2 (/A1), 4 (/A2), 6 (/A3), and 12 (/A4A)

Relay contact rating

250 VAC (50/60 Hz)/3 A, 250 VDC/0.1 A (for resistance load)

Output format

2 (/A1), 4 (/A2), 6 (/A3): NO-C-NC, and 12 (/A4A): NO-C

Relay operation

Energized/deenergized, AND/OR, hold/non-hold, and reflash settings are selectable.

RS-232 Interface (/C2) and RS-422A/485 Interface (/C3)

Item

Specifications

Connection

EIA RS-232(/C2) or EIA RS-422/485(/C3)

Protocol

Dedicated protocol or Modbus protocol

Synchronization

Start-stop synchronization

Transmission mode (RS-422A/485)

Four-wire half-duplex multi-drop connection (1:N (N = 1 to 32))

Data rate

1200, 2400, 4800, 9600, 19200, or 38400 bps

Data length

7 or 8 bits

Stop bit

1 bit

Parity

Odd, even, or none

Handshaking

Off:Off, XON:XON, XON:RS, and CS:RS

Communication distance (RS-422A/485)

1200 m

Modbus master

Reading information such as measured data from other instruments and writing this information to

registers

Modbus slave

Reading data from measurement and computation channels (/M1, /PM1, and /PWR1)

Reading and writing communication input data (/M1, /PM1, and /PWR1)

Some control commands such as memory start

Ethernet Communication Interface (/C7)

Item

Specifications

Electrical and mechanical specifications

Conforms to IEEE 802.3 (Ethernet frames conform to the DIX specification).

Medium

Ethernet (10BASE-T)

Protocol

Dedicated protocol as well as the TCP, IP, UDP, ICMP, ARP, DHCP, HTTP, FTP, SMTP, SNTP, and

Modbus protocols

E-mail client

Automatically send e-mail at specified times.

FTP client

Automatically transfer the following types of data files to the FTP server:

Display data, event data, snapshot data, and report data (/M1, /PM1, and /PWR1)

FTP Server

Moving and deleting files on the FX, managing directories, and generating file lists

Web server

Displaying the FX screen on a Web browser

SNTP client

Setting the FX time to the results of an SNTP server query

Cannot be used after hour 0 on January 1st, 2036.

SNTP server

Generating the FX’s time. Time resolution: 5 ms

Cannot be used after hour 0 on January 1st, 2036.

DHCP client

Automatically obtain the network address settings from the DHCP server.

Modbus client

Reads data from another device and writes to the registers.

Modbus server

Reading data from measurement and computation channels (/M1, /PM1, and /PWR1)

Reading and writing communication input data (/M1, /PM1, and /PWR1)

Some control commands such as memory start. Modbus client access limitations.

Setting/Measurement server Using the dedicated protocol, control and configure the FX, and generate data from the FX
Maintenance/test server

Outputs connection information and network information.

Instrument information server

Generating the information (such as the serial number and model name) of the connected FX

Advertising