Yokogawa Value Series FX1000 User Manual

Page 68

Advertising
background image

Common Operations

IM 04L21B01-01EN

2-9

3

2

1

4

5

6

7

8

9

10

11

12

13

14

App

Index

2.8 Initializing Settings and Clearing the Internal

Memory

Initialize the settings to default values. In addition, clear the data in the internal memory.
For the default settings, see appendix 4.

Setup Screen

Press

MENU (to switch to setting mode), hold down FUNC for 3 s (to switch to basic

setting mode), and select the

File/Initialize tab

1

>

Initialize.

1 On FXs that do not have a CF card slot or USB interface (/USB1 option), this is the Initialize

tab.

Setup Items

• Initialize

Settings

Description

Clear 1

Returns the basic setting mode and setting mode settings to their default values

and clears the data in the internal memory (measured data and computed data).

Clear 2

Returns the setting mode settings to their default values and clears the data in

the internal memory (measured data and computed data).

Clear 3

Clears the data in internal memory (measured data and computed data).

* Data in internal memory that is cleared: display data, event data, manual sampled data, report

data (/M1, /PM1, and /PWR1 options), and log information.

1.

Press the

Clear 1, Clear 2, or Clear 3 soft key.

2.

Press

DISP/ENTER.

A confirmation window opens.

3.

Select

Yes and press DISP/ENTER.

The specified operation is executed, and the FX returns to the operation mode.
If you do not want to initialize, select

No and press DISP/ENTER.

Advertising