Yokogawa Value Series FX1000 User Manual

Page 45

Advertising
background image

1-36

IM 04L21B01-01EN

Event Trigger Operation

When the event is set to output relay, internal switch, or alarm
If the output relay is activated, the internal switch is 1, or the alarm is occurring during

memory sampling, the event trigger is always activated. However, the number of times
the trigger is activated depends on the event data mode (single or repeat).

Resetting the Relative Timer

If the event is set to output relay, internal switch, match time timer, or alarm, the

resetting of the timer is not considered a timeout. (The action is not executed even if
the timer is used as an event.)

Loading the Setup

Can be specified as an action only when the event is set to remote control input.
Loads the setup data file, LOAD1.PDL, LOAD2.PDL, or LOAD3.PDL, in the root

directory of the CF card into the FX and updates the FX settings.

You must create a setup file and save it to the CF card in advance.

Time Adjustment

Time adjustment can be specified as an action only when the event is set to remote

control input. The internal clock of the FX is adjusted to the nearest hour through
remote control input.

• Operation When Memory Sampling Is Stopped

Difference from the Nearest Hour Operation
00 min 00 s to 01 min to 59 s

Truncates the minutes and seconds.

Example: 10 hours 01 min 50 s becomes 10 hours 00 min

00 s.

02 min 00 s to 57 min to 59 s

The time is not changed.

58 min 00 s to 59 min to 59 s

Rounds up the minutes and seconds.

Example: 10 hours 59 min 50 s becomes 11 hours 00 min

00 s.

• Operation during Memory Sampling (see section 1.10)

If the time difference Td between the time of the event and the nearest “on the

hour” hour is less than 2 minutes, the following operation is performed depending
on different conditions.
• If Td is less than or equal to the preset time difference, the clock is adjusted

gradually.

• If Td is greater than the preset time difference, the clock is adjusted immediately.

If Td is greater than or equal to 2 minutes, the clock is not adjusted.

1.6 Event Action and Remote Control Functions (/R1 and /PM1 Options)