9 performing calibration correction (/cc1 option) – Yokogawa Value Series FX1000 User Manual

Page 89

Advertising
background image

3-16

IM 04L21B01-01EN

3.9 Performing Calibration Correction (/CC1

Option)

The input value is calibrated using segments, and the result is used as a measured
value.

* You cannot specify calibration correction for a channel whose log scale (/LG1 option) mode

is set to LogType2.

For a description of the function, see section 1.1.

Setup Screen

Press

MENU (to switch to setting mode), and select the Menu tab > Meas channel >

Calibration correction.

Setup Items

True val

Number of set points

Calibration set points

Y5
Y4

Y3
Y2

Y1

MES val

X1 X2 X3 X4

X5

Condition

X1 < X2 < X3 < X4 < X5

• First-CH, Last-CH

Select the target channels. You can set consecutive channels whose range is set to

the same value as the first channel.

• Number of set points

Select the number of points that make up the segments (including the start and end

points) in the range of 2 to 16.

To disable calibration correction, select

Off.

• Number of set points > MES val, True val

Press the

Input soft key and enter the value.

For the MES value, set a value that is greater than the previous value.
Press the

Measure soft key to set the measured value at that point to MES val. If

you press the

Measure soft key when setting multiple channels simultaneously, the

measured value of the first channel are set to the

MES val of all channels.

Selectable Range of MES and True Values

• Channels on which linear scaling is specified

–30000 to 30000 (the decimal place is the same setting as the scale value)

• Other channels

Value in the measurable range of the selected range
Example: –2.0000 to 2.0000 for 2 V range

Note

• The calibration correction setting is set to Off if you change the

Mode or Range setting.

• Calibration correction cannot be specified on channels set to

Skip.

Advertising