App index – Yokogawa Value Series FX1000 User Manual

Page 228

Advertising
background image

10-3

IM 04L21B01-01EN

Troubleshooting

3

2

1

4

5

6

7

8

9

10

11

12

13

14

App

Index

Code

Message

Explanation/Countermeasures/Ref. section

91

Password is incorrect.

Enter the correct password. If you lost the password, it

must be reset.
See sections 8.2 and 8.3.

92

Press [ESC] key to change to the operation mode.

Press the ESC key.

93

String including space or all space cannot be specified.

Spaces are not allowed in the Web browser user name

and password.
See section 1.5 in the communication manual.

94

More than one address cannot be specified.

Only a single sender is allowed.
See section 1.4 in the communication manual.

95

Number entered exceeds channel number range.
Use another command.

Check the syntax of the Modbus command.
See sections 1.10 and 2.6 in the communication

manual.

96

This menu is locked.

Check the key lock setting. Release the key lock.
See section 8.1.

100

IP address doesn’t belong to class A, B, or C.

Check the IP address.
See section 1.3 in the communication manual.

101

The result of the masked IP address is all 0s or 1s.

Check the subnet mask.
See section 1.3 in the communication manual.

102

SUBNET mask is incorrect.

Check the subnet mask.
See section 1.3 in the communication manual.

103

The net part of default gateway is not equal to that of IP

address.

Check the IP address.
See section 1.3 in the communication manual.

105

This port number is already in use. Please enter a different

number.

Enter a different port number for each function.
See section 6.1 in the communication manual.

110

This user name is not registered.

Log in as a registered user.
See section 8.3.

113

Password entered is incorrect.

Enter the correct password.
See section 8.3.

116

This user name cannot be specified.

The user name is reserved.
See section 8.2.

119

This user name is unable to use this mode.

A user-level user cannot enter the Basic Setting mode
See section 8.2.

120

Measured value is incorrect. (in ascending order)

Set the calibration correction value to a value greater

than the previous value.
See section 3.9.

122

Measured value exceeds the range setting.

Check the channel input or the channel range setting.
See sections 3.3 and 3.9.

124

Password entry cannot be performed.

You cannot enter a password.

125

Character entry cannot be performed.

A display used to enter character screens is not being

shown on the FX.
See section 2.10.

126

You cannot use the same password.

Specify a different password.
See section 8.3.

127

Report kind overlaps and cannot be set up.

Change the overlapped report data type.
See section 9.5.

128

“Logout” cannot be set to “Hide”.

See sections 5.16 and 8.3.

10.1 A List of Messages

Advertising