2 alarms, 2 alarms -5, App index – Yokogawa Value Series FX1000 User Manual

Page 14: Alarm type

Advertising
background image

1-5

IM 04L21B01-01EN

Overview of Functions

3

2

1

4

5

6

7

8

9

10

11

12

13

14

App

Index

1.2 Alarms

This function generates an alarm when the measured data meets a certain condition. Up
to four alarms can be set for each channel.

Alarm Type

You can use the alarms shown below. The character inside the parentheses is the
symbol denoting each alarm.
For the alarm setting procedure, see section 3.7.

High Limit Alarm (H)

Alarm value

Alarm output ON

Alarm output ON

Alarm output ON

Alarm output ON

Alarm value

Hysteresis

Measured value

Measured value Hysteresis

Low Limit Alarm (L)

Alarm output ON

Alarm output ON

Delay time

Delay time

Alarm value

Alarm value

Measured value

Measured value

Delay High Limit Alarm (T)

Delay Low Limit Alarm (t)

High Limit on Rate-of-Change Alarm (R)

Low Limit on Rate-of-Change Alarm (r)

Alarm value

Alarm value

Hysteresis

Hysteresis

Difference in the measured values of two channels

Difference in the measured values of two channels

Difference High Limit Alarm (h)

Difference Low Limit Alarm (l)

Change in the

measured value

Change in the measured value

Measured value

Measured value

Time

Time

Interval (t2-t1)

Interval (t2-t1)

Amount of change in

the setting
|

|

T2

T1

t1

t2

T2

T1

t1

t2

T2-T1

Amount of change in

the setting

|

T2-T1

• Alarm Hysteresis

You can set a width (hysteresis) to the value used to activate and release alarms.
For the setting procedure, see section 3.5.

• Delay High Limit Alarm and Delay Low Limit Alarm

An alarm occurs when the measured value remains above or below the alarm value

for a specified time period (delay time).

• High Limit on Rate-of-Change Alarm and Low Limit on Rate-of-Change

Alarm

The rate-of-change of the measured values is checked over a certain time (interval).

An alarm occurs if the rate-of-change of the measured value in the rising/falling
direction is greater than or equal to the specified value.

The alarm value of the rate-of-change alarm is set using an absolute value. The

interval is derived using the following equation and set using the number of samples.

Interval = the scan interval × the number of samples

For the setting procedure, see section 3.5.

• Difference High Limit Alarm and Difference Low Limit Alarm

An alarm occurs when the difference in the measured values of two channels is

greater/less than or equal to the specified value. These alarms can be specified on
measurement channels set to difference computation.

Advertising