Extended input (/n3f), Remote control (/r1) – Yokogawa Value Series FX1000 User Manual

Page 277

Advertising
background image

13-12

IM 04L21B01-01EN

Extended Input (/N3F)

Item

Specifications

Measurement/display accuracy

Under standard operating conditions

Input Type

Measuring Range

Measurement Accuracy

Max.

Resolution

Kp vs Au7Fe

0.0 to 300.0 K

0 to 20 K

Within ±4.5 K

0.1 K

20 to 300 K

Within ±2.5 K

PLATINEL

0.0 to 1400.0°C

±(0.25% of rdg + 2.3°C)

0.1°C

PR40-20

0.0 to 1900.0°C

0 to 450°C

Accuracy not guaranteed

450 to 750°C

±(0.9% of rdg + 3.2°C)

750 to 1100°C

±(0.9% of rdg + 1.3°C)

1100 to 1900°C

±(0.9% of rdg + 0.4°C)

Thermocouple NiNiMo

0.0 to 1310.0°C

±(0.25% of rdg + 0.7°C)

W/WRe26

0.0 to 2400.0°C

0 to 400°C

±15.0°C

400 to 2400°C

±(0.2% of rdg + 2.0°C)

Type N (AWG14) 0.0 to 1300.0°C

±(0.2% of rdg + 1.3°C)

XK GOST

–200.0 to 600.0

–200 to –100°C

±(0.25% of rdg + 1.0°C)

–100 to 600°C

±(0.25% of rdg + 0.8°C)

Ni100 (SAMA)

–200.0 to 250.0°C ±(0.15% of rdg + 0.4°C)

Ni100 (DIN)

–60.0 to 180.0°C

±(0.15% of rdg + 0.4°C)

Ni120

–70.0 to 200.0°C

±(0.15% of rdg + 0.4°C)

RTD

1

Pt100 GOST

–200.0 to 600.0°C ±(0.15% of rdg + 0.3°C)

Cu100 GOST

–200.0 to 200.0°C ±(0.15% of rdg + 0.3°C)

Cu50 GOST

–200.0 to 200.0°C ±(0.4% of rdg + 0.5°C)

Pt200 (WEED)

–100.0 to 450.0°C ±(0.3% of rdg + 0.6°C)

1 Measuring current i = 1 mA

Input source resistance

Thermocouple input: 2 kΩ or less

RTD input:

1 Ω or less per wire (The resistance of all three wires must be equal).

Ambient temperature influence (with temperature variation of 10°C)
TC input

±(0.1% of rdg + 0.05% of range) or less, excluding the error of reference junction compensation

RTD input

±(0.2% of range + 2 digits) or less

Input source resistance
TC input

With variation of +1 kΩ: ±10 µV or less

RTD input

With variation of 1 Ω per wire (resistance of all three wires must be equal): ±(0.1% of rdg + 1 digit)

or less

With maximum difference of 100 mΩ between wires: Approx. 1 °C

Remote Control (/R1)

Item

Specifications

Number of input terminals

8

Input type

Isolated from the main circuitry through a photocoupler, built-in isolated power supply for the input

terminals, and shared common.

Input type and signal level
Voltage-free contact

Contact closed at 200 Ω or less and contact open at 100 kΩ or greater.

Open collector

ON voltage: 0.5 V or less (sink current 30 mA or more), leakage current when OFF: 0.25 mA or less

Allowable input voltage

5 VDC

Signal type

Level or edge (250 ms or more)

Action

Executes a specified action by applying a given signal to the remote signal input terminal.

Action assignment: Set using the event action function

13.5 Options