Setup examples, Setting the pseudo log range (logtype2) – Yokogawa Value Series FX1000 User Manual

Page 107

Advertising
background image

3-34

IM 04L21B01-01EN

Setup Examples

LogType2 [pseudo log] and LogType1 [nonlinear log] setup examples are shown below.

• Setting the Pseudo Log Range (LogType2)

This setting is for logging the pseudo log output (see the table below) of a vacuum gauge.

Pressure Reading [Pa] Output Voltage [V] Pressure Reading [Pa] Output Voltage [V]

1.30x10

-7

0.13

5.00x10

-4

3.50

5.00x10

-7

0.50

1.00x10

-3

4.10

1.00x10

-6

1.10

5.00x10

-3

4.50

5.00x10

-6

1.50

1.00x10

-2

5.10

1.00x10

-5

2.10

5.00x10

-2

5.50

5.00x10

-5

2.50

1.00x10

-1

6.10

1.00x10

-4

3.10

9.90x10

-1

6.99

1

Press

MENU (to switch to setting mode), hold down FUNC for 3 s (to switch to

basic setting mode), and select the

Environment tab > LOG input

Set

LogType2 under Input to Pseudo.

2

Press

MENU (to switch to setting mode), and select the Menu tab > Meas

channel > Range, Alarm.

Set the range as shown below.

Mode: LogType2
Range: 20V
Span: Lower 0.000, Upper 7.000
Scale: Lower 1.00E-7, Upper 1.00E0

Display Example

3.13 Using the Log Scale to Perform Measurements (/LG1 Option)

Advertising