5 optional terminal wiring, 5 optional terminal wiring -8, Warning – Yokogawa Value Series FX1000 User Manual

Page 257: Caution, Precautions to be taken while wiring, Wiring procedure

Advertising
background image

12-8

IM 04L21B01-01EN

12.5 Optional Terminal Wiring

WARNING

• To prevent electric shock while wiring, make sure that the power supply is turned

off.

• If a voltage of more than 30 VAC or 60 VDC is to be applied to the output

terminals, use ring-tongue crimp-on lugs with insulation sleeves on all terminals

to prevent the signal cables from slipping out when the screws become loose.

Furthermore, use double-insulated cables (dielectric strength of 3000 VAC or

more) for the signal cables on which a voltage of 30 VAC or 60 VDC or more is

to be applied. For all other signal cables, use basic insulated cables (dielectric

strength of 1500 VAC). To prevent electric shock, attach the terminal cover after

wiring and make sure not to touch the terminals.

CAUTION

• Use the following circuit voltages for the connection to the alarm/FAIL/status

output terminal.

• When the connection is to Mains Circuits (primary power supply circuits):

150 V or less

• When the connection is to circuits derived from Mains Circuits (secondary

power supply circuits): 250 V or less

(Keep the Mains Circuit voltage at 300 V or less, and use an isolation

transformer.)

• To prevent fire, use signal cables with a temperature rating of 70°C or more.

• Applying a strong tension to the input and output signal cables connected to

the FX may damage the cables or the FX terminals. To avoid applying tension

directly to the terminals, fix all cables to the rear of the mounting panel.

• Do not short the transmitter power supply output terminal or apply external

voltage to it. Doing so may damage the instrument.

• When using the transmitter power supply output terminal, do not use current that

is equal to or greater than the maximum output current (25 mADC). Doing so

may damage the instrument.

Note

For remote control wiring, use shielded wires to reduce noise. Connect the shield to the
functional ground terminal or the ground terminal of the FX.

Precautions to Be Taken While Wiring

We recommend that you use crimp-on lugs (designed for 3 mm screws) with

insulation sleeves to connect to the optional terminals.

Crimp-on lug (designed for 3 mm screws) with an insulation sleeve

Wiring Procedure

As shown in the figure on the next page, the optional terminal block is located on the rear

panel. The optional terminal block is only available when one of the following options,

which require I/O, is installed in the FX.

Alarm output relay (/A[ ] or /A4A), FAIL/status output relay (/F1), remote control

function (/R1), 24 V transmitter power supply output (/TPS[ ]), pulse input (/PM1),

power monitor (/PWR1), or communication interface (/C3)

Dedicated cover is attached to the optional terminal block.

1. Turn off the recorder and remove the terminal cover.

2. Connect the signal cables to the terminals. The recommended torque for tightening

the screws is 0.5 N•m.

3. Replace the terminal cover.

Advertising