3-36

IM 04L21B01-01EN

2

Configure calibration correction. Press

MENU (to switch to setting mode), and

select the

Menu tab > Meas channel > Calibration correction.

Set points for the calibration correction

Set the “MES val” values to the input scale values at the starting points of each line

and the “True val” values to the output scale values at the starting points of each line.

Number of

set points

MES

value [V]

True

value [V]

Number of

set points

MES

value [V]

True

value [V]

1

0.25

1.30

6

4.00

6.60

2

0.45

2.80

7

7.48

8.30

3

0.70

3.70

8

8.52

9.00

4

1.80

5.10

9

9.25

9.40

5

2.50

5.70

10

9.50

9.80

Note

For details on calibration correction (/CC1 option), see section 3.9.