3.11 Setting the Method of Detecting Over-Range

Values of Linearly Scaled Measurement

Channels

Setup Screen

Press

MENU (to switch to setting mode), hold down FUNC for 3 s (to switch to basic

setting mode), and select the

Environment tab > Input, Alarm

Setup Items

• Input > Value on over-range

Settings

Description

Free

The value is set to –over range if the value is less than –30000 and +over range

if the value is greater than 30000 excluding the decimal point. The value is

displayed as –Over and +Over, respectively.

Over

The value is set to –over range if the value is less than –5% of the scale and

+over range if the value is greater than 105%. The value is displayed as –Over

and +Over, respectively.

Example: If the scale is 0.0 to 200.0, the value is set to –over range if the value is

less than –10.0 of the scale and +over range if the value is greater than 210.0.

* The “Value on over-range” setting is not applied to channels that are set to log scale (/LG1

option). (See section 3.13.)

Note

For computations such as TLOG, CLOG, and report, the handling of the scale over-range value
can be set in advance.
See section 9.1.