Chapter 13 specifications, 1 signal input and alarms, 1 signal input and alarms -1 – Yokogawa Value Series FX1000 User Manual

Page 266: App index, Measurement input

Advertising
background image

13-1

IM 04L21B01-01EN

Specifications

3

2

1

4

5

6

7

8

9

10

11

12

13

14

App

Index

13.1 Signal Input and Alarms

Measurement Input

Item

Specifications

Number of inputs, scan interval, and A/D integration time

Model

No. of

Measurement Channels

Scan Interval

A/D converter integration

time

FX1002

2

125 ms, 250 ms

AUTO

1

, 50 Hz, 60 Hz

FX1004

4

FX1006

6

1 s, 2 s, 5 s

AUTO

1

, 50 Hz, 60 Hz,

100 ms

2

FX1008

8

FX1010

10

FX1012

12

1 AUTO: The FX automatically switches between 50 Hz and 60 Hz depending on the power

supply frequency.

2 You can only set the integration time to 100 ms on models FX1006 through FX1012. If you

set the integration time to 100 ms, you can only set the scan interval to 2 or 5 seconds.

Input Type

DC voltage, 1-5V, thermocouple (TC), resistance temperature detector (RTD), ON/OFF input (DI),

and DC current (by adding an external shut resistor)

Measurement range and measurable range

Input Type

Range

Measurable Range

DC voltage

20 mV

–20.000 to 20.000 mV

60 mV

–60.00 to 60.00 mV

200 mV

–200.00 to 200.00 mV

1 V

–1.0000 to 1.0000 V

2 V

–2.0000 to 2.0000 V

6 V

–6.000 to 6.000 V

20 V

–20.000 to 20.000 V

50 V

–50.00 to 50.00 V

1-5V

1 to 5 V

6

0.800 to 5.200 V

Thermocouple

R

1

0.0 to 1760.0°C

32.0 to 3200.0°F

S

1

0.0 to 1760.0°C

32.0 to 3200.0°F

B

1

0.0 to 1820.0°C

32.0 to 3308.0°F

K

1

–200.0 to 1370.0°C

–328.0 to 2498.0°F

E

1

–200.0 to 800.0°C

–328.0 to 1472.0°F

J

1

–200.0 to 1100.0°C

–328.0 to 2012.0°F

T

1

–200.0 to 400.0°C

–328.0 to 752.0°F

N

1

–270.0 to 1300.0°C

–454.0 to 2372.0°F

W

2

0.0 to 2315.0°C

32.0 to 4199.0°F

L

3

–200.0 to 900.0°C

–328.0 to 1652.0°F

U

3

–200.0 to 400.0°C

–328.0 to 752.0°F

WRe

4

0.0 to 2400.0°C

32.0 to 4352.0°F

RTD

Pt (Pt100)

5

–200.0 to 600.0°C

–328.0 to 1112.0°F

JPt(JPt100)

5

–200.0 to 550.0°C

–328.0 to 1022.0°F

DI

Level

0: Less than 2.4 V. 1: 2.4 V or higher (judged at the 6 V range)

Contact

7

0: Open. 1: Closed (parallel capacitance of 0.01 μF or less)

1 R, S, B, K, E, J, T, N: IEC584-1 (1995), DIN IEC584, JIS C1602-1995
2 W: W-5%Re/W-26%Re (Hoskins Mfg. Co.), ASTM E988
3 L: Fe–CuNi, DIN43710, U: Cu–CuNi, DIN43710
4 WRe: W-3%Re/W-25%Re (Hoskins Mfg. Co.)
5 Pt100: JIS C1604-1997, IEC751-1995, DIN IEC751-1996
JPt100: JIS C1604-1989, JIS C1606-1989. Measuring current i = 1 mA (Pt100, JPt100).
6 The range for linear scaling of 1-5V inputs. Burnout detection and low-cut functions are

available.

7 The detected current value is approximately 10 μA.

Chapter 13 Specifications

Advertising