3 data saving function, 3 data saving function -6, Configuration – Yokogawa Value Series FX1000 User Manual

Page 271: Data type, Display data and event data

Advertising
background image

13-6

IM 04L21B01-01EN

13.3 Data Saving Function

Configuration

Item

Specifications

Internal memory

Temporarily saves various types of data.

Medium

Flash memory

External storage medium (on FXs with a CF card slot)
Medium

CF card

Format

FAT32 or FAT16

Data Type

Item

Specifications

FX data types and file name extensions
Data Type

Extension

Notes

Display data

.DAD

Event data

.DAE

Manual sampled data

.DAM

Screen image data

.PNG

Setup data

.PDL

Report data

.DAR

/M1, /PM1, and /PWR1

Display Data and Event Data

Item

Specifications

Internal memory

File storage capacity

400 MB

Number of files

Up to 400

Operation

FIFO (First In First Out)

Display data

Target

Measurement/computation channel

Sampling intervals

Synchronized to the trend interval.

Content

Maximum or minimum value per sampling interval

Data size

Measurement channel data: 4 bytes/data value. Computation channel data: 8 bytes/data value.

File size

Up to 8 MB

Data format

Binary

Recording

Records data at all times.

Event data

Target

Measurement/computation channel.

Sampling interval

Determined by the sample rate.

125 ms, 250 ms, 500 ms, (FX1002 and FX1004 only)

1 s, 2 s, 5 s, 10 s, 30 s, 1 min, 2 min, 5 min, or 10 min

The sampling interval cannot be shorter than the scan interval.

Content

Data per sampling interval

Data size

Measurement channel data: 2 bytes/data value. Computation channel data: 4 bytes/data value.

File size

Up to 8 MB

Data format

Binary

Mode

Free: Records data at all times.

Trigger: Starts recording data when a certain event occurs and records for the specified interval.

Combinations of saved data Display data only, event data only, or display data and event data.
File size

See appendix 1.

Advertising