IM 04L21B01-01EN

• Creating a Display Data File or an Event Data File

This operation can be carried out when recording display data or when recording

event data in

Free mode. The file is created in the internal memory.

If

Auto save is set to On, the data file is saved to the CF card at the same time.

1.

In the operation mode, press

FUNC.

The Function menu appears.

2.

Press the

Save display or Save event soft key.

The display data or event data is saved to the CF card.

Time

Saved the previous time

Save operation

File

File

File

Saved this time

• Saving Measured Data Manually (Collectively Storing Unsaved Data)

This is the case for manually saving where

Auto save is set to Off (see section 6.2 for

details).

You can save to a CF card or to USB flash memory (/USB1 option). The procedure for

saving unsaved data to a CF card is described below.

For the procedure to save data to the USB flash memory, see section 2.11.

1.

Insert the CF card.

A confirmation window containing the message “There is data which is not saved to media.
Do you want to store to media?” appears.

2.

Select

Yes and press DISP/ENTER.

The unsaved data in internal memory will be saved to the CF card.

3.

Follow these steps to remove the CF card.

Press

FUNC (to display the Function menu) >

Media eject

soft key >

CF

soft key.

When the message “Media can be removed safely” appears, remove the CF card.

Note

• If there is not enough space on the external storage medium, the message “Not enough free

space on media” appears, and the data is not saved. If this message appears, replace the
external storage medium. Then, carry out the procedure again.

• You cannot abort the data save operation while it is in progress.