Measurement:meas<x>:stddev, Measurement:statistics:mode – GW Instek GDS-2000A series Programming Manual User Manual
Page 108

GDS-2000A Programming Manual
108
Example
:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:MEAS3:STATE ON
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:MINImum?
1.600E-02
Returns the minimum measurement result for
measurement number 3.
:MEASUrement:MEAS<X>:STDdev
Description
Returns the standard deviation for the selected
measurement from the last time the statistics were
reset. This is a statistics related command.
Syntax
:MEASUrement:MEAS<X>:STDdev?
Related
Commands
:MEASUrement:STATIstics:MODe
Parameter
MEAS<X> The automatic measurement number
from 1 to 8.
Example
:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:MEAS3:STATE ON
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:STDdev?
1.530E-03
Returns the standard deviation for measurement
number 3.
:MEASUrement:STATIstics:MODe
Description
Puts the statics measurement results on the display
or queries whether the statics are displayed.
Syntax
:MEASUrement:STATIstics:MODe {OFF | ON | ?}
Query
Set
Query