COMMAND DETAILS

51

Return Parameter ON

Channel is on.

OFF

Channel is off

Example

:CHANnel1:DISPlay ON
Turns on Channel 1

:CHANnel<X>:EXPand

Description

Sets Expand By Ground or Expand By Center for a
channel or queries its status.

Syntax

:CHANnel<X>:EXPand {GND | CENTer | ?}

Parameter

<X>

Channel 1,2,3,4

GND

Ground

CENTer Center

Return parameter GND

Expand By Ground

CENTER Expand By Center

Example

:CHANnel1:EXPand GND
Sets Channel 1 to Expand By Ground.

:CHANnel<X>:IMPedance?

Description

Returns the impedance of the oscilloscope.

Syntax

:CHANnel<X>:IMPedance?

Parameter

<x>

Channel

1/2/3/4 CH1/2/3/4

Return parameter <NR3> Returns the impedance value.

Example

:CHANnel1:IMPedance?
1.000000E+06
The impedance is 1M ohms.