Channel<x>:coupling, Channel<x>:deskew, Channel<x>:display – GW Instek GDS-2000A series Programming Manual User Manual

Page 50

Advertising
background image

GDS-2000A Programming Manual

50

:CHANnel<X>:COUPling

Description

Selects or returns the coupling mode.

Syntax

CHANnel<X>:COUPling {AC | DC | GND | ?}

Parameter

<X>

Channel 1,2,3,4

AC

AC coupling

DC

DC coupling

GND

Ground coupling

Return parameter Returns the coupling mode.

Example

:CHANnel1:COUPling DC
Sets the coupling to DC for Channel 1.

:CHANnel<X>:DESKew

Description

Sets the deskew time in seconds.

Syntax

:CHANnel<X>:DESKew { <NR3> | ?}

Parameter

<X>

Channel 1,2,3,4

<NR3> Deskew time: -5.00E -11 to 5.00E-11

-50ns to 50 ns.

Return parameter <NR3> Returns the deskew time.

Example

:CHANnel1:DESKew 1.300E-9
Sets the deskew time to 1.3 nano seconds.

:CHANnel<X>:DISPlay

Description

Turns a channel on/off or returns its status.

Syntax

:CHANnel<X>:DISPlay {OFF | ON | ?}

Parameter

<X>

Channel 1,2,3,4

OFF

Channel off

ON

Channel on

Set

Query

Set

Query

Set

Query

Advertising