Function<x>:amplitude, Function<x>:offset – GW Instek GDS-2000A series Programming Manual User Manual
Page 251

COMMAND DETAILS
251
Parameter
<x>
Module 1 or 2.
<NRf>
The frequency of the function
generator in Hz. (0.1 ~ 5000000)
Return parameter
Returns the frequency of the function generator
function in <NR3> format.
Example
:FUNCtion1:FREQuency 1000
Sets the function generator frequency to 1kHz for
the function generator in module slot 1.
:FUNCtion<x>:AMPlitude
Description
Sets or queries the function generator amplitude.
Syntax
:FUNCtion<x>:AMPlitude {<NRf>|?}
Related
commands
:FUNCtion<x>:MODe
Parameter
<x>
Module 1 or 2.
<NRf>
The amplitude of the waveform in
volts.
Return parameter
Returns the amplitude of the function generator
function in <NR3> format.
Example
:FUNCtion1:AMPlitude 2.00
Sets the function generator amplitude to 2.00 Vpp.
:FUNCtion<x>:OFFSet
Description
Sets or queries the function generator DC offset.
Syntax
:FUNCtion<x>:OFFSet {<NRf>|?}
Related
commands
:FUNCtion<x>:MODe
Parameter
<x>
Module 1 or 2.
<NRf>
(-1 ~ +1) The offset of the waveform
in volts.
Set
Query
Set
Query