Measure:fovshoot, Measure:fpreshoot – GW Instek GDS-2000A series Programming Manual User Manual
Page 83

COMMAND DETAILS
83
:MEASure:FOVShoot
Description
Returns the fall overshoot amplitude.
Syntax
:MEASure:FOVShoot{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter <NR3>
Returns the fall overshoot as a
percentage
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:FOVShoot?
1.27E+0
Selects Channel 1, and then measures the fall
overshoot.
:MEASure:FPReshoot
Description
Returns fall preshoot amplitude.
Syntax
:MEASure:FPReshoot{?}
Related
Commands
:MEASure:SOURce<X>
Returns
Returns the fall preshoot as <NR3>.
Return parameter <NR3>
Returns the fall preshoot as a percentage.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Query
Query