Trigger:bus:b1:uart:tx:data:size, Trigger:bus:b1:uart:tx:data:value – GW Instek GDS-2000A series Programming Manual User Manual
Page 141

COMMAND DETAILS
141
Return Parameter
Returns the data value.
Example1
:TRIGger:BUS:B1:UART:CONDition RXDATA
:TRIGger:BUS:B1:UART:RX:DATa:SIZe 1
:TRIGger:BUS:B1:UART:RX:DATa:VALue "1x1x0101"
Sets the value to 1x1x0101
Example 2
:TRIGger:BUS:B1:UART:RX:DATa:VALue?
1X1X0101
:TRIGger:BUS:B1:UART:TX:DATa:SIZe
Description
Sets or queries the number of bytes for UART data.
Note
This setting only applies when the UART trigger is
set to trigger on Tx Data
Syntax
:TRIGger:BUS:B1:UART:TX:DATa:SIZe {<NR1> | ?}
Related
commands
:TRIGger:BUS:B1:UART:CONDition
Parameter
<NR1>
Number of bytes (1 to 10).
Return parameter <NR1>
Returns the number of bytes.
Example
:TRIGger:BUS:B1:UART:TX:DATa:SIZe 5
Sets the number of bytes to 5.
:TRIGger:BUS:B1:UART:TX:DATa:VALue
Description
Sets or queries the triggering data value for the
UART bus when the bus is set to trigger on Tx
Data.
Syntax
:TRIGger:BUS:B1:UART:TX:DATa:VALue {string | ? }
Related
commands
:TRIGger:BUS:B1:UART:TX:DATa:SIZe
Set
Query
Set
Query