Measure:nedge, Measure:amplitude – GW Instek GDS-2000A series Programming Manual User Manual

Page 90

Advertising
background image

GDS-2000A Programming Manual

90

Note

Before using this command, select the
measurement channel. See the example below.

Example

:MEASure:SOURce1 CH1
:MEASure:PEDGE?
1.100E+01
Selects Channel 1, and then measures the number
of positive edges.

:MEASure:NEDGE

Description

Returns the number of negative edges.

Syntax

:MEASure:NEDGE{?}

Related
Commands

:MEASure:SOURce<X>

Return parameter <NR3>

Returns the number of negative edges.

Chan Off

Indicates the source channel is not
activated.

Note

Before using this command, select the
measurement channel. See the example below.

Example

:MEASure:SOURce1 CH1
:MEASure:NEDGE?
1.100E+01
Selects Channel 1, and then measures the number
of negative edges.

:MEASure:AMPlitude

Description

Returns the amplitude difference between the
Vhigh-Vlow.

Syntax

:MEASure:AMPlitude{?}

Related
Commands

:MEASure:SOURce<X>

Query

Query

Advertising