Trigger:bus:b1:can:identifier:direction – GW Instek GDS-2000A series Programming Manual User Manual
Page 152

GDS-2000A Programming Manual
152
Parameter/
Return parameter
<string>
The size of the string depends on the
data size setting. The string must be
enclosed in double quotes, “string”.
String contents:
x = don’t care
1 = binary 1
0 = binary 0
Example
:TRIGger:BUS:B1:CAN:CONDition ID
:TRIGger:BUS:B1:CAN:IDentifier:MODe STANDARD
:TRIGger:BUS:B1:CAN:IDentifier:VALue
"01100X1X01X"
:TRIGger:BUS:B1:CAN:IDentifier:VALue?
>01100X1X01X
:TRIGger:BUS:B1:CAN:IDentifier:DIRection
Description
Sets or queries the address bit as read, write or
don’t care.
Requirements: Firmware version V1.19 or later with
DS2-LA8 (or DS2-LA16).
Syntax
:TRIGger:BUS:B1:CAN:IDentifier:DIRection
{READ|WRITE|NOCARE|?}
Parameter/
Return parameter
READ
Sets read as the data direction
WRITE
Sets write as the data direction
NOCARE
Sets either as the data direction
Example1
:TRIGger:BUS:B1:CAN:IDentifier:DIRection?
>WRITE
Example2
:TRIGger:BUS:B1:CAN:IDentifier:DIRection READ
:TRIGger:BUS:B1:CAN:IDentifier:DIRection?
> READ
Set
Query