Search:trigger:bus:b1:can:identifier: mode, Search:trigger:bus:b1:can:identifier: value – GW Instek GDS-2000A series Programming Manual User Manual
Page 214

GDS-2000A Programming Manual
214
SEARCH:TRIGger:BUS:B1:CAN:IDentifier:
MODe
Description
Sets or returns the CAN addressing mode for the
bus.
Requirements: Firmware version V1.19 or later with
DS2-LA8 (or DS2-LA16).
Syntax
:SEARCH:TRIGger:BUS:B1:CAN:IDentifier:MODe
{STANDard|EXTended|?}
Parameter/
Return parameter
STANDard
Standard addressing mode
EXTended
Extended addressing mode
Example
:SEARCH:TRIGger:BUS:B1:CAN:IDentifier:MODe?
>STANDARD
:SEARCH:TRIGger:BUS:B1:CAN:IDentifier:MODe
EXTENDED
:SEARCH:TRIGger:BUS:B1:CAN:IDentifier:MODe?
>EXTENDED
:SEARCH:TRIGger:BUS:B1:CAN:IDentifier:
VALue
Description
Sets or returns the binary address string used for
the CAN search trigger.
Note: Only applicable when the search trigger
condition is set to ID or IDANDDATA.
Requirements: Firmware version V1.19 or later with
DS2-LA8 (or DS2-LA16).
Syntax
:SEARCH:TRIGger:BUS:B1:CAN:IDentifier:VALue
{<string>|?}
Related
Commands
:SEARCH:TRIGger:BUS:B1:CAN:IDentifier:MODe
Set
Query
Set
Query